| ±âº» ±¸¼º | C7000 Measurement cell, C7002 Cooling system, C 48 Oxygen station | 
  
    | Decomposition Vessel | C7010 | C7012 halogen resistant | 
      
        | Measuring Range max. | 30000 J | 
    
        | Temperature Measurement resolution | 0.0001 K | 
    
      | Oxygen operating pressure max. | 40 bar | 
    
      | Display | LED | 
    
      | Measuring mode | Double dry (ISO 1928) | 
    
      | Double dry reproducibility (1g benzoic acid NBS39i)
 | 0.2 % RSD | 
    
      | Working temperature max. | 30 ¡ÆC | 
    
      | Operation time double dry approx. | 3 min | 
    
      | Decomposition vessels | 8 | 
    
    | INTERFACES PC / Balance / Printer | RS 232 serial / RS 232 serial / Centronix | 
    
      | According to | ISO 1928, DIN CEN/TS 16023, DIN EN 15400, DIN EN 15170, DIN EN ISO 18125 | 
    
      | Type of cooling | Flow (C7002 Cooling system) | 
    
      | Cooling medium | Tap water (¼öµ¾¹°) | 
    
      | Cooling medium temperature min./max. | 12 ¡ÆC / 30 ¡ÆC | 
    
      | Cooling medium permissible operating pressure | 9 bar | 
    
      | Weight | 43 kg | 
    
      | Dimension (W¡¿D¡¿H) | 310¡¿490¡¿395 mm | 
      
        | Enviromental Conditions | Temp. 18 - 30 ¡ÆC / Humidity 80% | 
  
    | Power Supply | 220 - 240 V / 50/60 Hz / 260 W |